Kategorien
Noch keine Keywords zu diesem Inhalt eingetragen

ACS-CONTROL-SYSTEM GmbH Precont® PU4SE

You can find informationen about the smallest 4...20mA/HART pressure sensor Precont® PK4SH.
ACS-CONTROL-SYSTEM GmbH | www.acs-controlsystem.de | info@acs-controlsystem.de | Eggenfelden | Germany | 1 |
Precont® PU4SE
4...20mA
FSK-CPF9
RS485
Modbus
RTU
RS485
Pressure measurement Type:
In brief
Application
Your benets
Description
Pressure transmitter for general applications
Monitoring of relative pressure in gases, vapors,
liquids and dust
process
temperature
Protection
IP69K
pressure
up to
bar
600
Due to the device construction with
measuring ranges from 1 bar to 600
bar (gauge, adjustable), process
temperatures from –25°C to +100°C,
process materials Al2O3-ceramic /
CrNi-steel as well as the availability
of industrial standard process
connections like thread ISO 228-1, EN
837 manometer, thread ISO 228-1,
DIN EN ISO 1179-2 E or thread metric
resp. ANSI NPT the device is especially
suitable for the use for machinery and
plant engineering, air-conditioning
and refrigeration plant engineering,
hydraulic and pneumatic systems,
process industry or environmental
technology.
The pressure transmitter is suitable for
cost sensitive as well as demanding
measuring requirements.
Due to its high accuracy and the
digital adjustability by a FSK operating
unit or by RS485 Modbus RTU, the
device can be suited a wide variety of
applications.
The robust design and the high-quality
workmanship turns the device into
a very high quality product, which
even the most adverse environmental
conditions cannot aect, whether
the lowest temperatures when used
outdoors, extreme shock and vibration
stress or aggressive media.
A captive laser marking of the type
label ensures the identiability
throughout the entire lifetime of the
device.
Obviously is the optional marking of a
measurement point designation resp.
TAG, a customer label or of a neutral
type label, of course also per laser
marking, possible.
A LABS-free resp. silicone-free
version, a factory calibration with
calibration certicate and a customer
specic conguration resp. preset is
also optionally available like a material
test certicate EN10204 3.1.
Customer specic special versions
can be realized on request, e.g. other
connector, other process gasket
(EPDM, NBR, FFKM, etc.) or special
designs for the process connection.
General applications in
• Machinery and plant engineering
• Air-conditioning and refrigeration plant engineering
• Hydraulic and pneumatic systems
• Process industry
• Environmental technology
Wide range of applications
Measuring ranges from 1 bar up to 600 bar
Wide process temperature range –25°C to +100°C
High protection class IP69K
Wide environmental temperature range –25°C to +100°C
Ceramic internal diaphragm
High accuracy – characteristic deviation ≤ 0,5% of measuring range
Integrated evaluation electronic
Current output 4…20mA – HART® compliant (7.0)
Digital output and adjustability per RS485 Modbus RTU
Connector plug M12
fast
response time
| 2 | ACS-CONTROL-SYSTEM GmbH | www.acs-controlsystem.de | info@acs-controlsystem.de | Eggenfelden | Germany
Technical Data
Electrical connection
Electronic output – 2-wire, current 4…20mA
Conductor color standard connection cable M12
– A-coded:
BN = brown, BU = blue
For the HART® communication by a HART®
interface a minimum communication resistance of
250Ω has to be taken into account.
Electronic output – 4-wire, RS485
Conductor color standard connection cable M12
– A-coded:
BN = brown, WH = white, BU = blue, BK = black
Technical Data
Supply voltage: 9…35VDC, reverse polarity protected
Supply current: ≤ 22mA Electronic output type A – 2-wire, current 4…20mA
≤ 10mA Electronic output type V – 4-wire, RS485 Modbus RTU
RS485 Modbus RTU
Interface RS485, bidirectional
Signal Digital – Modbus RTU
Address 001 (001…247)
Transmission rate 9600 Baud (4800 / 9600 / 19200 / 38400)
Parity Odd (None / Odd / Even)
Step response time T90 ≤ 5ms (td = 0s)
Start-up time tOn ≤ 0,1s
Analogue output 4...20mA
Operating range: 3,9…21mA, min. 3,8mA, max. 22mA
Permitted load: ≤ (US - 9V) / 22mA
Start-up time: ≤ 0,1s
Communication FSK modulated current signal – HART® compliant (7.0)
Signal 0,5mASS – 1200Hz / 2200Hz
Communication resistor ≥ 250Ω, external
Activity tOn ≤ 30s / td ≥ 0,1s
Address 0 (0…15)
Transmission rate 1200 Bit/s
Measuring accuracy
Characteristic deviation: ≤ ± 0,5% FS
Long term drift: ≤ ± 0,2% FS / year not cumulative
Temperature deviation ≤ ±0,5% FS / 10K
Materials
Diaphragm: (process wetted) Ceramic aluminum oxide Al2O3 – 96%
Process connection: (process wetted) Steel 1.4404/316L / Steel 1.4571/316Ti
Terminal enclosure: CrNi-steel
Gaskets: (process wetted) FPM – uorelastomere (Viton®)
Environmental conditions
Environmental temperature: – 25°C...+100°C
Process temperature: – 25°C...+100°C
Process pressure: 0…1 bar / ≤ 4 bar; 0…4 bar / ≤ 12,5 bar; 0…10 bar / ≤ 25 bar;
0…40 bar / ≤ 125 bar; 0…100 bar / ≤ 200 bar; 0…600 bar / ≤ 800 bar
Protection: IP69K/IP67 (EN/IEC 60529)
ACS-CONTROL-SYSTEM GmbH | www.acs-controlsystem.de | info@acs-controlsystem.de | Eggenfelden | Germany | 3 |
Precont® PU4SE
Dimension drawings Type:
Type 1 – Thread ISO 228-1 – G½” – EN 837 Type 3 – Thread ISO 228-1 – G¼“ –
DIN EN ISO 1179-2 E
| 4 | ACS-CONTROL-SYSTEM GmbH | www.acs-controlsystem.de | info@acs-controlsystem.de | Eggenfelden | Germany
Stand 10/2016
Precont® C R
PU4S VE S SS
Order code
Type
PU4S Standard
Measuring system – material diaphragm (process wetted) / sensor type
E Ceramic 96%, DMS
Approval
S Standard
Process connection
1 Thread ISO 228-1 – G½”B, EN 837 manometer
3 Thread ISO 228-1 – G¼”B, , DIN EN ISO 1179-2 E
Y others
Material gaskets (process wetted)
1 FPM – uorelastomere (e.g. Viton®)
Y others
Material process connection (process wetted)
V CrNi-steel
Material terminal enclosure
C CrNi-steel
Measuring range
05 0…1 bar
08 0…4 bar
10 0…10 bar
13 0…40 bar
19 0…100 bar
24 0…600 bar
YY Special measuring range
Electronic – output
A 4-wire, current 4…20mA, HART® compliant
V 4-wire, RS485, Modbus RTU
Electronic – function
S Standard
Process temperature
0 Standard –25°C…+100°C
Y others
Pressure type
R Gauge pressure
Measuring system – accuracy
4 0,5%
Y others
Electrical connection
S Plug M12x1